Spectrograph camera Specs:(Note; under construction, Jeff Hickey Oct. 5, 2008)

Slit 1 x 30 arcsecond slit

Wavelength coverage 1.0 - 2.4 micron

Spectral resolution 2500-2700

Sampling ~ 2.7 pixels per resolution element

Minimum integration time 4 seconds

Gain ~ 3.8 e-/DN

Saturation level ~ 28000 DN (conservatively)

Data sampling CDS, Fowler, or sample-up-the ramp.

Read Noise 10 e- (CDS), 5 e- (8 samples)

3.5 e- (16 samples)

Dark Current ~ 0.085 e-/sec

Time to background limit ~ 300 seconds



Status: Guider/Imager camera: Observatory Instrument

Guider/Imager Specs:

Wavelength 1-2.5 micron

Size of Field 4' X 4'

Scale .234"/pixel

Minimum integration time 1.8 seconds

MCT Engineering Grade Hawaii II CCD 1024 X 1024

Data sampling CDS, CCD linear to 30,000 ADU, however,
in differencing mode continuously exposed, the displayed image goes non-linear ~18,000 ADU

Read Noise <10 e- (CDS)