|
LRIS-B Overview |
|
Experienced LRIS users should find it very straightforward to use LRIS-B, since the new spectrograph channel behaves in most ways very similarly to the red side. Even those interested primarily in wavelengths longward of 5000 Å may find that the blue side is very useful for improving observing efficiency (particularly for slit mask setups) and for allowing the use of higher dispersion gratings without sacrificing spectral coverage.
The blue spectrograph has completely independent filters, dispersing elements, and spectrograph camera from those on the red side. The blue camera was designed to produce very close to the same image scale at the detector as the red camera, so that with the new science grade CCD mosaic on the blue side, the pixel scale is 0.015/0.024 times that on the red side, or about 0.135"/pix. The total slit length seen by the blue camera incorporates the entire unvignetted field of the instrument, so that the full 8.0' field is useable on the blue side (as opposed to ~7.3 arc minutes on the red side with the current detector). Each of the dichroic, grism, and filter systems consist of two main subsystems:
The two sides share the same off-axis paraboloidal collimator mirror; this collimator mirror has been coated with a special hybrid coating (developed at LLNL) that reflects as efficiently as Ag above 4000 Angstroms and better than Al in the UV and blue.
At present, for spectrscopy using dichroics, there are ghost spectra that have been determined to be second order light dispersed by the red side grating that is returned to the dichroic (and which passes through because the dichroic is basically a long-pass filter) and then makes it onto the blue side detector as zeroth order light (i.e., it is not further dispersed by the blue grism). The ghosting will thus depend upon red-side grating tilt and by the prism angle of the grism in use. We have ordered a short-pass filter for use in the blue side filter carousel that will block these ghosts for many setups (the ghost line intensity is about 0.5% of the red side line intensity--they are easily seen in arcs but can be quite subtle in taken on the sky)--it is expected to be available by August 2002. In the meantime, if your observations require deep near-UV and blue spectroscopy, I would recommend using the dichroic mirror, which eliminates this ghosting issue. There are no obvious ghosts seen with the 1200 l/mm grism-- because the large prism angle diverts the zero-th order image of the dispersed red light away from the camera. Ghost images should not be a problem for any imaging observations.
It is possible to shift between spectroscopic and imaging mode simply by removing or inserting the grism into the beam (choosing deploy or stow on the XLRIS GUI). This can be done in less than 30 seconds, and hence we anticipate that most users whose programs involve multislit spectroscopy will want to do their slit mask alignments using the LRIS-B imaging mode, leaving LRIS-R in spectroscopic mode (with grating in place). This time savings can amount to as much as 35-45 minutes of observing time during a typical LRIS multi-slit night.
Grisms, like gratings, produce "second-order" light. For extremely blue sources, there may be significant second-order light appearing longward of (first-order) 6000 A for the 300 and 400-line grisms now that the UV QE of the CCDs is very high.
Note that on LRIS-B one has the option of pulling any filter out of the beam without selecting a different filter (simply choose "clear" from the blue filter list on the XLRIS GUI to stow the filter in the carousel).
The most efficient means of doing slit mask alignments will be to obtain field images unfiltered using only the dichroic to define the bandpass, so that deploying the grism is the only change to the spectrograph configuration necessary to go from alignment to the start of the spectroscopic integration.
For example, it is possible to image in (u' or B) and (R or I) simultaneously with the right choice of dichroic -- D560 would probably be the best choice in this case. As another example, one can choose to image in V and I simultaneously, with the D680 dichroic in place. There are some issues that remain to be ironed out concerning the best way to focus the telescope for imaging, depending on whether the observer cares more about the blue or red side images (only the red side can currently be used for running MALIGN). Obviously, it is possible to make ``bad'' choices of filter and dichroic combinations (e.g., D560 would be a bad choice for V band imaging on either side, and D680 would be a bad choice for R band imaging on the red side) and at the moment such mistakes are not caught by the instrument control software.
For spectroscopic observations, there are many combinations. The choice will depend upon what resolution is needed in the blue versus red, etc. Some examples appear in the accompanying table.
| Grism | Dichroic | Grating |
|---|---|---|
| 400/3400 | D560 | 600/5000 [coverage ~3150-8000 Å] 400/8500 [coverage ~3150-9400 Å] |
| 1200/3400 | D460 | [any red grating; gap in spectral coverage] |
| 300/5000 | D680 | 832/8200 [coverage ~3200-8600 Å] |
| 600/4000 | D500 | 600/5000 600/7500 |
Since many new optical components were installed inside the LRIS spectrograph body, there is the possibility of unwanted internal reflections or scattered light being introduced. While every effort was made to add baffles in several places, it is possible that the internal baffling can be improved. If you find evidence for scattered light or internal reflection problems, please alert your instrument specialist and provide as much information (images, ideally) as you can so that improvements might be made.
There is now a very large number of moving mechanisms inside LRIS; all 4 mechanisms on the blue side (focus control, grism, filter, dichroic carousels and transports) can move simultaneously, while 2 mechanisms (grating plus slitmask/filter/focus) can move simultaneously on the red side. All communications with the motor control is done via terminal servers (one for the red side and one for the blue side) and occasionally you might find that a move timed out or was left in an unknown state. We have found that any failures of blue side mechanisms can often be fixed simply by trying the failed move again. Until we have had more experience running both sides of the spectrograph simultaneously it will be necessary for observers to be on their toes to make sure the spectrograph is configured in the desired manner.
Insertion of a dichroic does not alter the LRIS-R camera focus value, but it will have a small effect on position of the beam footprint at the LRIS-R grating (which results in small shifts of the wavelength range seen at the CCD). The dichroics should be essentially identical in terms of these shifts.
Because it is a reflection off of the dichroic that sends light into LRIS-B, and each dichroic is independently adjustable in tilt and piston, there are small focus differences between the different dichroics (and the focus offsets should be redetermined whenever adjustments are made to the dichroics in their cells) which are tabulated for your use. In addition, the footprint at the CCD will vary somewhat from dichroic to dichroic. Thus, if you will be using images of slit masks taken in the afternoon to aid in slit mask alignments during the night, you should be sure that you are using the same dichroic as you will use for your setup images during the night.